Procedure - Solid Samples

Press the Analyze button on the instrument screen again, the loading head slide-block will close and the lower electrode will open.


Clean the upper and lower electrode manually, or, if applicable, remove the crucible and press the analyze button to clean with the automatic cleaner


Add approximately 0.05 g of *** Graphite Powder to a *** Graphite Crucible.


Firmly place the crucible on the lower electrode tip or appropriate autoloader position.


Press the Analyze button on the instrument screen, the lower electrode will close and the analysis sequence will start and end automatically



上海禹重给您提供各种氧氮氢分析仪,欢迎来电咨询!北京元素分析仪厂家供应

北京元素分析仪厂家供应,分析仪

Procedure – Powder/Chip Samples

Instrument calibration/drift correction.

Press the Analyze button on the instrument screen. After a short delay, the loading head slide-block will open.

Place the Nickel Capsule containing the sample into the open port at the top of the loading head.

Press the Analyze button on the instrument screen again, the loading head slide-block will close and the lower electrode will open.

Clean the upper and lower electrode either manually or remove the crucible and press the analyze button to clean with an automatic cleaner if applicable.



北京元素分析仪厂家供应装有自动旁路阀的可快速拆装氧水捕集装置。

北京元素分析仪厂家供应,分析仪

User-Friendly Brand Software Cornerstone

LECO's exclusive brand software enables the user to have complete access to analysis Cornerstone control, method settings, diagnostics, reporting, and more in a highly organized and immersive environment. Designed through a collaboration of customer feedback and innovative engineering,Cornerstone allows the user to conduct all of their day-to-day operations within a single Analysis screen designed for speed and ease-of-use. Our innovative grouping of sample data into sets and replicates simplifies the data output and automatically calculates relevant statistics, alleviating the need for additional data processing.


Procedure - Solid Samples

Determine the instrument blank.

Repeat steps 2b through 2g a minimum of three times.


Set the blank following the procedure outlined in the operator's instruction manual.


Instrument calibration/drift correction.

Login a minimum of 3 Standard replicates.


Weigh approximately *** grams of a calibration/drift standard, enter the mass and standard identification into appropriate replicate fields.



Note: LECO Reference Materia***o not require preparation.See preparation statement on the reference material certificate.





工具条、滑动条和下拉式菜单简化了校正和数据处理的参数设定。

北京元素分析仪厂家供应,分析仪

Procedure – Powder/Chip Samples

Prepare the instrument as outlined in the operator's instruction manual.


Determine the instrument blank.

Login a minimum of three Blank reps.

Press the Analyze button on the instrument screen. After a short delay, the loading head slide-block will open. Insert a *** Nickel Capsule (leave capsule open) into the open port at top of loading head.

Note: samples using automation should be placed in the appropriate autoloader position before starting the analysis sequence. Once the sequence has started, the automatic analysis will start and end automatically.


20位试样自动进样系统。北京元素分析仪厂家供应

Cornerstone氢氧氮分析软件同时支持***的书记文档格式和可变的报告格式。北京元素分析仪厂家供应

Summary


The determination of the amount of oxygen, nitrogen, and hydrogen in iron, steel, nickel-, and cobalt-base alloys represents some of the most important quality metrics for these materials. Oxygen is used to create steel from pig iron by removing excess carbon. Oxygen content must be controlled to limit the amount of carbon monoxide that can be formed during solidification which may cause excessive porosity. Nitrogen is considered both an impurity as well as an important alloying agent. Itcan be present as a nitride or interstitially in its gaseous form. Increased nitrogen content is known to increase yield and tensile strength, thus decreasing ductility and formability. Excessive levels may evolve during solidification thus increasing porosity. High hydrogen content is the primary cause of embrittlement, blistering and flaking due to its high

mobility through the lattice and provides no potential alloying benefits. The ONH836 utilizes a high-power

electrode furnace to quickly and efficiently release the target gases from within the sample, which allows

for a very rapid simultaneous determination of oxygen, nitrogen, and hydrogen.


北京元素分析仪厂家供应